If you need to cool an equipment, you can use a simple formula to calculate how much heat your device will generate, then according to the calculated results , you can choose the most suitable product model of our thermostatic bath instrument.
Eg. Cooling water inlet temperature=15℃
Cooling water inlet temperature=16.2℃
Cooling water flow=6L/min
So the heat generated by the application system can be calculated according to the following formula:
P=m/t ×g × c ×ΔT
Among them, m/t (cooling water flow) 360L/hour
g (proportion of water) 1.0kg/L
c ( specific heat of water) 1.0Kcal/kg/k
ΔT (Cooling water inlet temperature difference) 1.2k
P=360×1.0×1.0×1.2= 432Kcal/hr = 502W (note: 860 Kcal/hr =1,000 W)
Click here to see more about thermo circulation bath in BuenoBiotech