Downloads

How to Connect the External Circulation of Thermo Circulating Bath

1. Add liquid medium in the external circulation thermostatic water bath | thermostatic water bath circulator , and the liquid level of the liquid medium shall not be lower than 30mm on the working table.

1. Add liquid medium in the external circulation thermostatic water bath | thermostatic water bath circulator, and the liquid level of the liquid medium shall not be lower than 30mm on the working table.

2. Selection of liquid medium:
A. At 5℃ to 85℃, pure water is generally used as the liquid medium.
B. When the working temperature is 85℃ or above, Methyl silicone oil  is generally used as the liquid medium.
C. When the operating temperature is above 95℃, the selected oil's flash point value should be higher than the operating temperature above 15℃.
Note: operating temperature here refers to the temperature to be reached by the liquid medium in the tank

3.Connection of circulating pump:

A. Connection of internal circulation pump, connect the outlet pipe with the hose of the inlet pipe (with a hose randomly).

B. The external circulation pump is connected to the external circulation. The hose of the outlet pipe is connected to the inlet of the tank external container, and the inlet pipe is connected to the outlet of the tank external container.

external circulation connection photoQQ图片20190704153628

Click here to see more about Thermostatic Circulating Water Bath in BuneoBiotech

 

4. When using thermostatic circulating bath, such as the working temperature >100℃, it is recommended that the external connection of circulating pump should adopt metal pipe.


5. Plug in and turn on the "main power" switch.

 

External circulation thermostatic water bath | thermostatic water bath circulator in BuenoBiotech

 

 